Clinical evaluation for SaMD and AI: where Notified Bodies push back
For software as a medical device (SaMD) and AI-enabled medical software, clinical evaluation is where Notified Bodies most often raise deficiencies. Until release R007, our open-source baseline only had a single cross-reference in SOP-007 and no procedure behind it. That gap is now closed with SOP-025 (Clinical Evaluation and Post-Market Clinical Follow-up), built on EU MDR Article 61 and Annex XIV, plus Swiss MedDO Articles 21(4) and 46(3).
Writing a clinical evaluation procedure for software meant addressing the recurring mistakes that hold up real conformity assessments.
1. Dataset metrics are not clinical data
Software teams naturally start with what they already have: sensitivity, specificity, or AUC measured on a retrospective dataset. Under MDCG 2020-1 and IMDRF N41, software clinical evaluation has four distinct links: valid clinical association, technical (analytical) performance, clinical performance, and clinical benefit. Algorithm testing on stored data belongs to technical performance. Presenting it as "clinical data" under MDR Article 2(48) draws an immediate finding.
SOP-025 keeps those layers separate and sets concrete rules for AI performance studies: how the reference standard (ground truth) was established, whether raters were blinded to the device output, and how disagreement was resolved; strict patient and, where relevant, site separation between development and a fixed test set; and how performance holds across equipment, subgroups, and real-world prevalence in the intended care setting. Acceptance criteria are also locked through a two-step Clinical Evaluation Plan approval, after the state-of-the-art review and before device data are appraised, so thresholds cannot be adjusted after seeing the results.
2. "Zero complaints" is not proof of safety
Complaints and vigilance reports count as clinical data under MDR Article 2(48), which tempts teams with fielded or legacy software to treat a quiet support inbox as clinical evidence. MDCG 2020-6 Appendix III says plainly that complaints and vigilance data sit at Rank 7, suffer from under-reporting bias, and are never sufficient on their own to demonstrate GSPR conformity.
SOP-025 (§6.7(4)) and our YAML appraisal template cap complaint and vigilance data at "supportive" weight. They are useful for spotting safety trends and must be combined with appraised clinical data or proactive PMCF, never used as standalone proof of safety or performance. Where a manufacturer argues that clinical data are not appropriate under Article 61(10), SOP-025 requires a four-part justification and states explicitly that Article 61(10) does not remove the Article 61(4) clinical investigation requirement for Class III or implantable devices.
3. Notifying the Notified Body on a model change is not enough
Software changes faster than annual CER updates. Without a screening step, model retraining, threshold adjustments, or input and output changes reach production while the CER still describes an older version. SOP-025 (§6.13) screens every software change raised under change control (SOP-009) into one of four documented outcomes: no clinical effect (with rationale), a CER addendum, new analytical or clinical evidence before release, or a full CER update. Changes to a model, decision threshold, input type, output, or intended user always require at least a CER addendum and written reasoning if new evidence is not collected.
A review comment during our own PR caught a distinction many procedures miss: for changes covered by MDR Annex IX §4.10 or Annex X §5.2, informing the Notified Body is not enough. Where Notified Body approval is required, the changed device must not be released until that approval is obtained, linked in the change record, and checked in the final release decision under SOP-020.
4. Clinical evaluation fails in a silo
A clinical evaluation SOP that sits apart from engineering and post-market work decays quickly. In R007, SOP-025 is wired into the Quality Manual (QM-001) and seven existing procedures: medical device file content (SOP-007), design validation under ISO 13485 7.3.7 (SOP-008), change impact screening (SOP-009), PMS and PMCF planning with CER review at least at PSUR cadence under Article 86(1) (SOP-014), benefit-risk alignment with residual risk under ISO 14971 (SOP-018), release readiness (SOP-020), and management review inputs (SOP-004).
Ten templates under records/clinical/ support the procedure: evaluator declaration (MEDDEV 2.7/1 rev. 4 §6.4), CEP, literature search protocol, state-of-the-art assessment, YAML clinical data appraisal, equivalence assessment (including Article 61(5) MDR compliance and data access), Article 61(10) justification, a CER structured on the MDCG 2020-13 CEAR topics, PMCF plan (MDCG 2020-7), and PMCF evaluation report (MDCG 2020-8).
5. You don't need GitHub to use this: give it to your agents
As I wrote in my response to OpenRegulatory, much of today's regulatory drafting and gap-checking runs through AI agents. Even if your company operates a traditional eQMS and has no plans to move to Git, the clinical evaluation layer (SOP-025, all ten templates, and the clause-level MDR clinical evaluation gap analysis) is plain text and YAML, indexed for agents at qms.dearauditor.ch/llms.txt.
Point your agent at SOP-025 and the gap analysis, give it your existing clinical evaluation SOP or draft CER, and ask where your procedure misses an Annex XIV requirement, an Article 7 claims check, or a pre-release change gate. Qualified clinical evaluators still own the clinical judgment and sign the record, while the agent handles the first-pass cross-check in minutes.
What remains yours
The usual boundary holds: no upstream template proves clinical benefit for your device, replaces qualified evaluators, or runs a clinical investigation under ISO 14155. ISO 18969 is still at FDIS stage and stays tracked as an open review item once published. Everything in R007 is anchored to the consolidated EU MDR text, Swiss MedDO, MDCG guidance, MEDDEV 2.7/1 rev. 4, and IMDRF documents, with no reproduced standard or third-party white-paper text.
You can read the procedure, templates, and gap matrix in github.com/AliakseiT/dearauditor-qms-baseline (release QMS-2026-11-24-R007). If your clinical or regulatory review finds a gap, open an issue. That is how the pre-release Notified Body approval gate made it in before we tagged R007.